Biomet Microfixation, Inc. Lactosorb Pectus Stabilizer, model number: 01-3805. Distributed and Manufactured by : Biomet Microfixation, Jacksonville, FL. LactoSorb Pectus Stabilizer is a surgical implant intended to aid in the treatment of Pectus Excavatum deformities. The LactoSorb Pectus Stabilizer provides the surgeon with a means to secure the Pectus Support Bar to the chest wall. The Pectus Support Bar is used to reposition bony structures (sternum, breastbone) by applying internal force outwardly eliminating the fun Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Lactosorb Pectus Stabilizer, model number: 01-3805. Distributed and Manufactured by : Biomet Microfixation, Jacksonville, FL. LactoSorb Pectus Stabilizer is a surgical implant intended to aid in the treatment of Pectus Excavatum deformities. The LactoSorb Pectus Stabilizer provides the surgeon with a means to secure the Pectus Support Bar to the chest wall. The Pectus Support Bar is used to reposition bony structures (sternum, breastbone) by applying internal force outwardly eliminating the fun
Brand
Biomet Microfixation, Inc.
Lot Codes / Batch Numbers
All lots.
Products Sold
All lots.
Biomet Microfixation, Inc. is recalling Lactosorb Pectus Stabilizer, model number: 01-3805. Distributed and Manufactured by : Biomet Microfi due to This recall is based on the occurrence of known adverse effects identified in the device instructions for use. Bending, fracture, loosening, rubbing . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
This recall is based on the occurrence of known adverse effects identified in the device instructions for use. Bending, fracture, loosening, rubbing and migration of the devices can occur as a result of excessive activity, trauma or load bearing. Implantation of foreign materials can result in an inflammatory response or allergic reaction.
Recommended Action
Per FDA guidance
An "Urgent Medical Device Recall Notice" was issued via certified letter or email on March 20, 2009 to consignees. Customers were instructed to immediately remove and return the affected in-stock product to Biomet Microfixation, Inc. Customers were also asked to complete and return the enclosed Inventory Reconciliation sheet via fax (904-741-9425) to the firm, and to notify recipients of any recalled products that had been further distributed. For questions about the recall, contact Biomet Microfixation at 1-800-874-7711 or 904-741-4400, extension 468, Monday through Friday, 8am-4pm EST.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MI, MN, MO, NE, NH, NJ, NM, NY, NC, OH, OK, PA, TX, VA, WI, DC