BioSense Webster, Inc. Circuito Interior Norte, # 1820 Parque Industrial Salvarcar Cd. Juarez, Chihuahua Mexico Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02
Brand
BioSense Webster, Inc. Circuito Interior Norte, # 1820 Parque Industrial Salvarcar Cd. Juarez, Chihuahua Mexico
Lot Codes / Batch Numbers
All lot numbers
Products Sold
All lot numbers
BioSense Webster, Inc. Circuito Interior Norte, # 1820 Parque Industrial Salvarcar Cd. Juarez, Chihuahua Mexico is recalling Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257- due to Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT Catheters, which may cause the ablation catheter icon to shift on the map viewer when RF energy is being applied. Therefore the user, attempting a linear ablation in the location displayed on the CARTO RMT screen, may ultimately apply RF energy to u
Recommended Action
Per FDA guidance
The firm sent an "Urgent Medical Device Correction" letter to consignees on October 6, 2008 informing consignees. Users were advised to pay "close attention to the location of the ablation catheter icon on the map viewer when applying RF energy." If consignees suspect that they have a defective NAVISTAR RMT Catheter, they were instructed to notify Biosense Webster and to return the defective catheter for evaluation and refund/replacement. Customer were instructed to contact their local Biosense Webster representative for any questions or further information. No return reply was requested.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, CT, FL, GA, IL, IN, IA, KS, KY, MA, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI