Biosense Webster Inc Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e., recording and stimulation only. The device is designed to obtain electrograms in the atrial regions of the heart. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e., recording and stimulation only. The device is designed to obtain electrograms in the atrial regions of the heart.
Brand
Biosense Webster Inc
Lot Codes / Batch Numbers
All lots
Products Sold
All lots
Biosense Webster Inc is recalling Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The cathet due to This recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral va. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
This recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral valve, requiring surgical procedures, valve may be torn while trying to dislodge, unable to retract the catheter through the sheath and require surgical intervention to remove, pulling on catheter to remove may lead to large atrial septal tear, and sep
Recommended Action
Per FDA guidance
A letter entitled, "Urgent Safety Field Notice" dated March 25, 2008, and certification form were sent to all customers and Johnson & Johnson affiliates who purchased the product in the past year as of March 25, 2008, due to the expiration date. For customers in the US, two letters and forms were sent out - one addressed to the EP/Cardiology Lab and a second to the Risk Management Office. If you have questions or have not received a letter contact Customer Service at 1-800-729-9010.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.