VARIPULSE Ablation Catheter (Biosense) – Neurovascular Events (2025)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recording) and, when used for TRUPULSE Generator, for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation. The catheter provides location information when used with the CARTO 3 System.
Brand
Biosense Webster, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Product Number: D141201-12 UDI-DI code: 10846835025460 Batch Numbers: 31483574L 31483560L 31483581L 31483542L 31487313L 31483552L 31483544L 31487311L 31483550L 31483558L 31483575L 31483556L 31487312L 31483553L 31483565L 31483576L 31483547L 31487320L 31483549L 31483561L 31483548L 31483564L 31483541L 31483546L 31483562L 31483566L 31483583L 31483567L
Biosense Webster, Inc. is recalling VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use due to Due to an observed trend of neurovascular events. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to an observed trend of neurovascular events
Recommended Action
Per FDA guidance
On 01/05/2025, the firm called and emailed customers to inform them to pause the US External Evaluations and all Varipulse cases due an observed trend in neurovascular events associated with the use of the VariPulse Bi-Directional Ablation Catheters distributed between December 2024 and January 2025. For U.S. Customers, Biosense Webster/JNJ Medtech EP is pausing the US External Evaluation and all US Varipulse cases. Healthcare providers who have treated patients using the products prior to the pause should continue to follow those patients according to their standard of care. For any questions, contact Biosense Webster/JNJ MEDTECH EP representatives On 01/08/2025, the firm released the information on the firm's parent company's website https://www.jnj.com/media-center/press-releases/johnson-johnson-statement-on-varipulse-u-s-external-evaluation
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AR, CA, CO, KS, LA, MA, MO, NY, OH, TX