Boston Scientific Corporation Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.
Brand
Boston Scientific Corporation
Lot Codes / Batch Numbers
Serial Numbers: 008885, 010823, 010615, 007886
Products Sold
Serial Numbers: 008885, 010823, 010615, 007886
Boston Scientific Corporation is recalling Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LAT due to Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update A. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
Recommended Action
Per FDA guidance
The firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.