Boston Scientific Corporation EMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator, Model A219, sterile. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
EMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator, Model A219, sterile.
Brand
Boston Scientific Corporation
Lot Codes / Batch Numbers
UPN: 00802526584404, Serial numbers: 169280, 169298, 169494, 169753, 169776, 169934, 169941, 170003, 169547, 169151, 169315, 169514, 169851, 169879, 169292, 169696, 169786, 169788, 169852, 170002, 169400, 169405, 169898, 169973, 169885, 169335, 169233, 169608, 169304, 169410, 166452, 169278, 169319, 169321, 169651, 169688, 169864, 169920, 169235, 169277, 169834, 169472, 169417, 169202, 169329, 169426, 169511, 169529, 169630, 169757, 169963, 167092, 169236, 169301, 169310, 169316, 169393, 169436, 169485, 169537, 169543, 169565, 169567, 169571, 169579, 169582, 169599, 169601, 169606, 169663, 169771, 169238, 169323, 169680, 169270, 169265, 169372. UPN: 00802526581519, Serial numbers 169312, 169409, 169421, 169428, 169521, 169561, 169493, 169654, 169653, 169635, 169731, 169477, 169495, 169496, 169474, 169491, 169468, 169455, 169442, 169446, 169435, 169447, 168979, 169271, 169384, 169353, 169371, 169351, 169203, 169469, 169471, 169314, 169342, 169475, 169448, 169404, 169407, 169450, 169396, 169470, 169356, 169373, 169402, 169413, 169441, 169539, 169545, 169551, 169674, 169267, 169350, 169589, 169307, 169595, 169596, 169600, 169286, 169603, 169605, 169206, 169340, 169502, 169505, 169510, 169348.
Products Sold
UPN: 00802526584404, Serial numbers: 169280, 169298, 169494, 169753, 169776, 169934, 169941, 170003, 169547, 169151, 169315, 169514, 169851, 169879, 169292, 169696, 169786, 169788, 169852, 170002, 169400, 169405, 169898, 169973, 169885, 169335, 169233, 169608, 169304, 169410, 166452, 169278, 169319, 169321, 169651, 169688, 169864, 169920, 169235, 169277, 169834, 169472, 169417, 169202, 169329, 169426, 169511, 169529, 169630, 169757, 169963, 167092, 169236, 169301, 169310, 169316, 169393, 169436, 169485, 169537, 169543, 169565, 169567, 169571, 169579, 169582, 169599, 169601, 169606, 169663, 169771, 169238, 169323, 169680, 169270, 169265, 169372. UPN: 00802526581519, Serial numbers 169312, 169409, 169421, 169428, 169521, 169561, 169493, 169654, 169653, 169635, 169731, 169477, 169495, 169496, 169474, 169491, 169468, 169455, 169442, 169446, 169435, 169447, 168979, 169271, 169384, 169353, 169371, 169351, 169203, 169469, 169471, 169314, 169342, 169475, 169448, 169404, 169407, 169450, 169396, 169470, 169356, 169373, 169402, 169413, 169441, 169539, 169545, 169551, 169674, 169267, 169350, 169589, 169307, 169595, 169596, 169600, 169286, 169603, 169605, 169206, 169340, 169502, 169505, 169510, 169348.
Boston Scientific Corporation is recalling EMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator, Model A219, sterile. due to There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
Recommended Action
Per FDA guidance
The recalling firm emailed Dear Health Care Professional or Hospital Administrator retrieval letters dated 9/30/2022 to their sales representatives on 9/30/2022 who began hand delivering them immediately in the U.S. for delivery to those consignees who were believed to have unimplanted devices in their possession. The letter described the issue and said their Boston Scientific sales representative is available to support this retrieval and listed the following steps to be taken: (1) Immediately segregate all affected product listed in Appendix A of the letter to ensure it will not be used; (2) Complete the attached Reply Verification Tracking Form and return it via the email address provided or via FAX; and (3) Ship product back to Boston Scientific, and the address was provided. OUS notification began 10/3/2022. For the U.S. patients who have already been implanted with the device, correction letters are to be hand-delivered to the doctors the week of 10/17/2022 and the sales representatives will send confirmation of delivery back to the recalling firm. No reply form was enclosed with the correction letter. These letters will be personalized/customized with the name of the following or implanting physicians or clinics. The letters will be dated October 2022 and contain the patient's name, date of birth, and the model/serial number and GTIN implanted. The letter will describe the issue, clinical impact, and recommendations regarding follow-up interval, patient notes entry, and medial records. Prophylactic replacement is not recommended as the issue does not affect the S-ICDs performance or the actual battery longevity.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026