Extractor Pro Retrieval Catheter (Boston Scientific) – Incorrect Labeling (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);
Brand
Boston Scientific Corporation
Lot Codes / Batch Numbers
Instructions for Use: 51878205, Pre-Printed Ouch: 51368417, 1. UPN (REF): M00547030 (9-12mm Below), UDI-DI: 08714729790280, Lot Numbers: 36398057, 2. UPN (REF): (1) M00547000(9-12mm Above), UDI-DI: 08714729790266, Lot Numbers: 36426495
Products Sold
Instructions for Use: 51878205; Pre-Printed Ouch: 51368417; 1. UPN (REF): M00547030 (9-12mm Below); UDI-DI: 08714729790280; Lot Numbers: 36398057; 2. UPN (REF): (1) M00547000(9-12mm Above); UDI-DI: 08714729790266; Lot Numbers: 36426495;
Boston Scientific Corporation is recalling Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 due to The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
Recommended Action
Per FDA guidance
On August 28, 2025 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Instructions: 1. Immediately cease further distribution or use of any remaining affected devices. Segregate the units in a secure place until they can be returned. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Return affected devices.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.