Boston Scientific Corporation UpsylonTM Y Mesh Kit with Colpassist Vaginal Positioning Device - Product Usage: intended for use as a bridging material for sacrocolposuspension/sacrocolpopexy (laparotomy, laparoscopic or robotic approach Material No.M0068318220 Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. Note: The UpsylonTM Y Mesh Kit consists of UpsylonTM Y Mesh and ColpassistTM Vaginal Positioning Device packaged and sterilized separately and then combined into a kit carton. Thi Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
UpsylonTM Y Mesh Kit with Colpassist Vaginal Positioning Device - Product Usage: intended for use as a bridging material for sacrocolposuspension/sacrocolpopexy (laparotomy, laparoscopic or robotic approach Material No.M0068318220 Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. Note: The UpsylonTM Y Mesh Kit consists of UpsylonTM Y Mesh and ColpassistTM Vaginal Positioning Device packaged and sterilized separately and then combined into a kit carton. Thi
Brand
Boston Scientific Corporation
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
All unexpired lots: C003726 C003737 C003739 C003752 C003767 C003778 C003782 C003791 C003814 C003818 C003831 C003845 C003846 C003874 C003875 C003898 C003899 C003921 C003942 C003954 C003970 C003988 C003989 C003999 C004000 C004011 C004012 C004031 C004032 C004033 C004048 C004049 C004084 C004085 C004086 C004097 C004098 C004099 C004120 C004132 C004143 C004150 C004159 C004175 C004176 C004196 C004205 C004213 C004218 C004223 GTIN: 08714729848103
Boston Scientific Corporation is recalling UpsylonTM Y Mesh Kit with Colpassist Vaginal Positioning Device - Product Usage: intended for use as due to Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operat. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection
Recommended Action
Per FDA guidance
Boston Scientific issued Urgent Medical Device Product Removal on 4/15/21 via FedX. Letter states reason for recall, health risk and action to take; Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. This removal affects only the UPNs and batches listed below and in Attachment 1. No other UPNs or batches are impacted by this removal. Please cease further distribution or use of any remaining product affected by this removal immediately. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions and indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. BSC is notifying regulatory authorities of this removal as required. Please read carefully through the removal instructions included with this notification. Your local Sales Representative can answer any questions that you may have regarding this notification.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.