Boston Scientific Corporation WATCHDOG Hemostasis Valve Kit, UPN H74939343A022 (outer package), UPN H74939343A020 (inner package) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
WATCHDOG Hemostasis Valve Kit, UPN H74939343A022 (outer package), UPN H74939343A020 (inner package)
Brand
Boston Scientific Corporation
Lot Codes / Batch Numbers
GTIN 08714729996019 (outer package), GTIN 08714729965794 (inner package), Batch numbers: 30272850, 30542947, 30661041, 30845826, 30935565
Products Sold
GTIN 08714729996019 (outer package), GTIN 08714729965794 (inner package), Batch numbers: 30272850,30542947,30661041,30845826, 30935565
Boston Scientific Corporation is recalling WATCHDOG Hemostasis Valve Kit, UPN H74939343A022 (outer package), UPN H74939343A020 (inner package) due to The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.
Recommended Action
Per FDA guidance
Boston Scientific issued an URGENT MEDICAL DEVICE PRODUCT REMOVAL notice on 04/05/2023 by overnight mail. The notice explained the issue and the hazard and requested the consignee segregate the devices pending return. Medical facilities who distributed the devices and distributors were directed to send the notice to their customers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026