Boston Scientific Scimed Titanium Greenfield Vena Cava Filter with 12 Fr./4,0 mm Introducer System. Catalog no. M001503010. Sterile EO. Single Use Only. Manufacturer: Boston Scientific Cork Ltd., Business and Technology Park, Model Farm Rd., Cork, Ireland. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. The devices are permanently implanted devices designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. It is Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Titanium Greenfield Vena Cava Filter with 12 Fr./4,0 mm Introducer System. Catalog no. M001503010. Sterile EO. Single Use Only. Manufacturer: Boston Scientific Cork Ltd., Business and Technology Park, Model Farm Rd., Cork, Ireland. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. The devices are permanently implanted devices designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. It is
Brand
Boston Scientific Scimed
Lot Codes / Batch Numbers
Lot numbers: 7129578 and 7124116
Products Sold
Lot numbers: 7129578 and 7124116
Boston Scientific Scimed is recalling Titanium Greenfield Vena Cava Filter with 12 Fr./4,0 mm Introducer System. Catalog no. M001503010. due to Certain units within eleven lots of the 12 Fr Greenfield Vena Cava Filter may lack a taper on the braided sheath of the delivery system. The lack of . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Certain units within eleven lots of the 12 Fr Greenfield Vena Cava Filter may lack a taper on the braided sheath of the delivery system. The lack of a taper could lead to vessel damage as a result of the sheath catching on the edge of the vein and causing a tear.
Recommended Action
Per FDA guidance
Urgent Recall Notice dated 08/04/05 was Federal Express''d to customers for delivery on 08/05/05. This notice informs customers of the issue and identifies affected product. Customers are asked to immediately discontinue use of and segregate these recalled lots. Steps for Voluntary Recall and Tracking/Verification Form were sent to the Risk Manager only. The Tracking/Verification Form is to be completed by the Risk Manager even if there are no affected lots to return. To return affected product, the customer must call Boston Scientific Corporation (BSC) to obtain a Returned Goods Authorization (RGA) number. Returned Goods are to be packaged and sent to BSC with the Tracking/Verification Form and using the enclosed red/white field action return shipping label to: Boston Scientific Marina Bay, 500 Commander Shea Blvd., Quincy, MA 02171
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution