Breg Inc, An Orthofix Company Power Supply Accessory (model AD-0660G) Power supply accessory is used with the Polar Care Kodiak unit, a motorized cold-therapy device. The Power Supply is plugged into a standard electrical wall socket and provides low voltage power to the Kodiak unit. The power supply is not used with other Polar Care models such as PC300 and PC500. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Power Supply Accessory (model AD-0660G) Power supply accessory is used with the Polar Care Kodiak unit, a motorized cold-therapy device. The Power Supply is plugged into a standard electrical wall socket and provides low voltage power to the Kodiak unit. The power supply is not used with other Polar Care models such as PC300 and PC500.
Brand
Breg Inc, An Orthofix Company
Lot Codes / Batch Numbers
Lots 3508, 3908, 4308, and 4808.
Products Sold
Lots 3508, 3908, 4308, and 4808.
Breg Inc, An Orthofix Company is recalling Power Supply Accessory (model AD-0660G) Power supply accessory is used with the Polar Care Kodiak u due to Power supplies short-circuited at the electrical outlet upon initial use of the power supply, causing the circuit breaker for the electrical outlet to. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Power supplies short-circuited at the electrical outlet upon initial use of the power supply, causing the circuit breaker for the electrical outlet to trip.
Recommended Action
Per FDA guidance
An Urgent: Medical Device Recall Notification was sent to customers on June 18, 2009. FOR USERS: Customers are advised to immediately open all unused Kodiak cold therapy units and inspect the lot code of the Power Supply included with the Kodiak unit. If the lot code is 3508, 3908, 4308, or 4808, then remove the power supply, cut the cord using scissors and discard. Call Breg Customer Service for a replacement: 1-800-321-0607. Complete and return the response form ASAP. FOR DISTRIBUTORS: Send Breg your distribution information, including contact and quantities. Breg will notify your customers of the recall directly.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026