Caire, Inc. CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit
Brand
Caire, Inc.
Lot Codes / Batch Numbers
UDI/DI M766133374030, Serial Numbers: CBB3022300368, CBB3022300390, CBB3022300470, CBB3022300472, CBB3022300479
Products Sold
UDI/DI M766133374030, Serial Numbers: CBB3022300368, CBB3022300390, CBB3022300470, CBB3022300472, CBB3022300479
Caire, Inc. is recalling CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit due to An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
Recommended Action
Per FDA guidance
The firm issued an URGENT: MEDICAL DEVICE RECALL notice its consignees be email and telephone. The notice explained the issue and requested the removal of the affected products from the field and return to CAIRE. CAIRE will immediately replace the above serial numbers with new Liberator devices. Please contact CAIRE Customer Service via phone at 1-800-482-2473 or 770-721-7700 or via email at customerservice.usa@caireinc.com to arrange for the return and replacement of the above serial numbers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.