Cardiac Science Corporation Powerheart AED G3 Automated External Defibrillator Model 9300A (Automatic). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Powerheart AED G3 Automated External Defibrillator Model 9300A (Automatic). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.
Brand
Cardiac Science Corporation
Lot Codes / Batch Numbers
serial numbers 4000002, 4000004, 4000006, 4000007, 4000009, 4000010, 4000011, 4000013, 4000014, 4000016, 4000018, 4000019, 4000021 thru 4000024, 4000030, 4000031, 4000034, 4000040, 4000044, 4000045, 4000047, 4000048, 4000050, 4000051, 4000054, 4000057, 4000060, 4000063, 4000064, 4000065, 4000067, 4000072, 4000075, 4000079, 4000080, 4000082, 4000085, 4000086, 4000089, 4000090, 4000095, 4000097, 4000098, 4000099, 4000136, 4000138, 4000140, 4000145, 4000156, 4000162, 4000164, 4000179, 4000184, 4000186, 4000187, 4000192, 4000209, 4000215, 4000227, 4000229, 4000230, 4000232, 4000247, 4000249, 4000250, 4000255, 4000256, 4000264, 4000271, 4000274, 4000287, 4000288, 4000291, 4000295, 4000305, 4000306, 4000314, 4000322, 4000323, 4000324, 4000326, 4000330 thru 4000335, 4000337, 4000342.
Products Sold
serial numbers 4000002, 4000004, 4000006, 4000007, 4000009, 4000010, 4000011, 4000013, 4000014, 4000016, 4000018, 4000019, 4000021 thru 4000024, 4000030, 4000031, 4000034, 4000040, 4000044, 4000045, 4000047, 4000048, 4000050, 4000051, 4000054, 4000057, 4000060, 4000063, 4000064, 4000065, 4000067, 4000072, 4000075, 4000079, 4000080, 4000082, 4000085, 4000086, 4000089, 4000090, 4000095, 4000097, 4000098, 4000099, 4000136, 4000138, 4000140, 4000145, 4000156, 4000162, 4000164, 4000179, 4000184, 4000186, 4000187, 4000192, 4000209, 4000215, 4000227, 4000229, 4000230, 4000232, 4000247, 4000249, 4000250, 4000255, 4000256, 4000264, 4000271, 4000274, 4000287, 4000288, 4000291, 4000295, 4000305, 4000306, 4000314, 4000322, 4000323, 4000324, 4000326, 4000330 thru 4000335, 4000337, 4000342.
Cardiac Science Corporation is recalling Powerheart AED G3 Automated External Defibrillator Model 9300A (Automatic). Cardiac Science AEDs ar due to Some Powerheart AED Model 9300A may contain a suspect electronic component and there is a remote likelihood that the function of this device may be in. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Some Powerheart AED Model 9300A may contain a suspect electronic component and there is a remote likelihood that the function of this device may be interrupted during the charge.
Recommended Action
Per FDA guidance
Affected consignees were notified via telephone. If no phone contact could be made, a Urgent: Medical Device Recall letter dated 02/02/06 was sent with return receipt requested. Customers were asked to locate affected devices and remove them from service. A new AED would be sent to the customers and the customers are asked to return the affected device.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.