Cardiac Science Corporation Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).
Brand
Cardiac Science Corporation
Lot Codes / Batch Numbers
Lot/Serial Numbers: 11040077414, 11040077801, 11040077859, 11040077861, 11040077969, 11040077977, 11040077974, 11040078366, 11040078367, 11040078375, 11040078376, 11040078379, 11040078385, 11040078384, 11040078368, 11040078395, 11040078452, 11040078891, 11040079014, 11040079161, 11040079568, 11040079567, 11040079569, 11040079671, 11040079734, 11040079939, 11040079880, 11040080124
Products Sold
Lot/Serial Numbers: 11040077414 ,11040077801 ,11040077859 ,11040077861 ,11040077969 ,11040077977 ,11040077974 ,11040078366 ,11040078367 ,11040078375 ,11040078376 ,11040078379 ,11040078385 ,11040078384 ,11040078368 ,11040078395 ,11040078452 ,11040078891 ,11040079014 ,11040079161 ,11040079568 ,11040079567 ,11040079569 ,11040079671 ,11040079734 ,11040079939 ,11040079880 ,11040080124
Cardiac Science Corporation is recalling Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science). due to 28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
Recommended Action
Per FDA guidance
On April 3, 2009 the recalling firms notified consignees by certified letter to domestic customers and via FedEx to international customers with delivery confirmation signature required. Customers were advised to remove treadmill from service until it is tested by a Field Service Engineer. For customers with questions or those that have not already scheduled a service call with Cardiac Science Field Service Engineers, please contact the firm at 1.800.426.0337, option 1.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution