Cardiac Science, Inc. Powerheart AED G3 Automated External Defibrillator Model 9300E. Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Powerheart AED G3 Automated External Defibrillator Model 9300E. Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.
Brand
Cardiac Science, Inc.
Lot Codes / Batch Numbers
Model no. 9300E serial numbers: 360814 thru 361239, 361320 thru 362359, 362387 thru 362393, 362396 thru 362713, 362916 thru 364194, 364283 thru 364514, and 364606 thru 364914.
Products Sold
Model no. 9300E serial numbers: 360814 thru 361239, 361320 thru 362359, 362387 thru 362393, 362396 thru 362713, 362916 thru 364194, 364283 thru 364514, and 364606 thru 364914.
Cardiac Science, Inc. is recalling Powerheart AED G3 Automated External Defibrillator Model 9300E. Cardiac Science AEDs are manufactur due to AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may no. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the event that charge noise is presrent it will delay the first defibrillation shock in the 3 shock sequence.
Recommended Action
Per FDA guidance
Cardiac Science will contact customers via telephone to set up a time for Clinical Educator to visit the customer and conduct a field test on each AED. AEDs that fail the field test will be replaced either on the spot or within 24 hours. AEDs failing the test are either removed by the Clinical Educator or are returned to Cardiac Science upon receiving the new AED. Cardiac Science anticipates testing to be done by the end of December 2004.