Cardinal Health 200, LLC Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777
Brand
Cardinal Health 200, LLC
Lot Codes / Batch Numbers
Model Number/Product Code: 1180100777. UDI/DI: 10192253034530 each, 20192253034537 box, 50192253034538 case. Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601.
Products Sold
Model Number/Product Code: 1180100777. UDI/DI: 10192253034530 each, 20192253034537 box, 50192253034538 case. Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601.
Cardinal Health 200, LLC is recalling Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777 due to Syringe contains conical tip that is not compatible with certain needleless IV connector(s).. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Syringe contains conical tip that is not compatible with certain needleless IV connector(s).
Recommended Action
Per FDA guidance
Cardinal Health notified consignees on about 09/13/2023 via letter titled "URGENT MEDICAL DEVICE PRODUCT CORRECTION." Customers were instructed to review inventory for the affected product code and lot, ensure all personnel are informed cease utilizing the affected product with needleless IV connectors, post a copy of the notification in the storeroom where the product is stored, and notify any customers if the product has been further distributed/forwarded. Additionally, customers were asked to complete and return the included acknowledgment form.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026