Cardinal Health 303 dba Cardinal Health The Alaris PC Unit, Model 8015 (manufactured or serviced between April 2008 and January 2009). The Alaris PC Unit is the central programming and power supply component for the Alaris System modules. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
The Alaris PC Unit, Model 8015 (manufactured or serviced between April 2008 and January 2009). The Alaris PC Unit is the central programming and power supply component for the Alaris System modules.
Brand
Cardinal Health 303 dba Cardinal Health
Lot Codes / Batch Numbers
Model 8015 (manufactured or serviced between April 2008 and January 2009).
Products Sold
Model 8015 (manufactured or serviced between April 2008 and January 2009).
Cardinal Health 303 dba Cardinal Health is recalling The Alaris PC Unit, Model 8015 (manufactured or serviced between April 2008 and January 2009). The due to There is a potential in the specified ESD (electrostatic discharge) protection circuitry in the keypad of Alaris PC units manufactured or serviced bet. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential in the specified ESD (electrostatic discharge) protection circuitry in the keypad of Alaris PC units manufactured or serviced between April 2008 through January 2009. In these units there is a potential for the keypad not being responsive, key entries without key presses, or key entries that register incorrectly which may result in an over-infusion or under-infusion. An over
Recommended Action
Per FDA guidance
The recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PROBLEM #2 Syringe volume Warning Message for Alaris PCA module; (C) PROBLEM #3 Electrostatic Discharge (ESD) protection circuitry; (D) PROBLEM #4 Fluid Ingress and (E) PROBLEM #5 Alaris System Inter-Unit Interface (IUI) Connectors Labeling. The recall notice included "Required Action for Users & Follow-up actions by Cardinal Health for all five problems. Customers were also asked to report any adverse reactions experienced with the use of this product, and/or quality problems to the FDA's MedWatch Program. In addition, customer were provided the following contact information for the recall: (1) Cardinal Health Recall Center for general questions: By phone at 1-888-562-6018 7am to 5pm (Pacific) ; (2) Customer Advocacy for adverse reports: By phone at 1-800-854-7120, Option 1, Option 1, Option 3 or by email at customerfeedback@cardinalhealth.com 24 hours a day, Sunday-Saturday; (3) Technical Support for Technical Questions regarding the Alaris System: By phone at 1-888-562-6018, 6am-5pm (Pacific). The firm followed-up with the release of a nationwide press statement regarding this issue on July 29, 2009. The press release indicated serial numbers of affected devices, as well as CareFusions short term instructions to customers, and the firms strategy to fix the affected devices can be found at: www.cardinalhealth.com/alaris/medical-device-recall/ (This informatio
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution