Cardinal Health 303 Inc DBA Alaris Products U9 Socket Integrated Circuit ("U9 IC") on the display board of the following Alaris System modules built on or before October 4, 2005: (1) Alaris Syringe module (Model 8110) (a.k.a. Medley Syringe module) (2) Alaris PCA module (Model 8120) (a.k.a. Medley PCA module), CardinalHealth, San Diego, CA 92130 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
U9 Socket Integrated Circuit ("U9 IC") on the display board of the following Alaris System modules built on or before October 4, 2005: (1) Alaris Syringe module (Model 8110) (a.k.a. Medley Syringe module) (2) Alaris PCA module (Model 8120) (a.k.a. Medley PCA module), CardinalHealth, San Diego, CA 92130
Brand
Cardinal Health 303 Inc DBA Alaris Products
Lot Codes / Batch Numbers
Pending
Products Sold
Pending
Cardinal Health 303 Inc DBA Alaris Products is recalling U9 Socket Integrated Circuit ("U9 IC") on the display board of the following Alaris System modules b due to A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the Alaris Syringe Module ("Syringe module") and Alaris PCA module ("PCA module") for modules manufactured on or before October 4, 2005. The user will be required to re
Recommended Action
Per FDA guidance
An Urgent Medical Device Recall Notification letter was sent to consignees on April 10, 2008. Four customer letters were issued. The U19 IC issue affects only the Pump module and the U9 IC issue affects only the Syringe and PCA modules. Therefore there will be one letter for customers affected by all three issues Ul9 IC, U9 IC, and the IUI, one letter for affected Ul9 IC and U9 IC customers, one letter for U19 IC-only customers and one letter for U9 IC-only customers. All mailings were conducted through consignees (distributors as well as direct customers for the affected devices). Distributors will be sent multiple copies of the Recall Notification and Customer Response Cards. The Recall letters all includeRequired Actions sections which give their customers specific information about the issue and what the customers are to do and includes a web site link for each specific problem. Each recall letter also has chart of Cardinal Health contacts support.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, WY