BD Pyxis MedBank Tower (CareFusion) – Label Printing Error (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9
Brand
CareFusion 303, Inc.
Lot Codes / Batch Numbers
UDI: 10885403512568/ Serial Number: 15684798
Products Sold
UDI: 10885403512568/ Serial Number: 15684798
CareFusion 303, Inc. is recalling BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9 due to Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorre. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
Recommended Action
Per FDA guidance
On July 11, 2024, CareFusion, subsidiary of Becton, Dickinson and Company (BD) issued a "Urgent Medical Device Correction" Notification, via mail. On June 24, 2024, BD issued Customer Release Notes via their firm's customer portal. BD ask consignees to take the following actions: Clinicians: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. Users should verify the contents of the bin location being restocked. 2. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Pyxis" MedBank" CUBIE" Replenishment Station devices to. 3. Please promptly complete and return the enclosed Customer Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter Distributor: 1. Please provide a copy of the attached customer letter to all facilities within your distribution network that have any product listed in Appendix 1. 2. Please promptly complete and return the enclosed Distributor Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.