Centurion Medical Products Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM110. Aspiration tray containing Portex uncuffed pediatric sized tracheal tubes. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM110. Aspiration tray containing Portex uncuffed pediatric sized tracheal tubes.
Brand
Centurion Medical Products
Lot Codes / Batch Numbers
Lots 901087, 903267, 907307, 910297, 901218, 908258, 2009020990 and 2009022390.
Products Sold
Lots 901087, 903267, 907307, 910297, 901218, 908258, 2009020990 and 2009022390.
Centurion Medical Products is recalling Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State due to The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete obstruction of the airway.
Recommended Action
Per FDA guidance
The firm's sales representative was notified of the recall on September 22, 2009 via email and the sole customer was notified by letter dated September 23, 2009. All product was removed from the customer account on September 28, 2009 and returned to the recalling firm.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IL