Xpert SA Nasal Complete (Cepheid) – test result error (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique
Brand
Cepheid
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Lot Numbers: 27815 27816 Batch Numbers: 1001424179 1001424181
Cepheid is recalling Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Comp due to As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.
Recommended Action
Per FDA guidance
On 08/21/2025, the firm sent via FedEx an "Urgent Medical Device Recall" to customers to inform customers that Cepheid conducted a routine post-market stability study for GXSACOMP-10 kit batch 1001424179 lot 27815, which failed to meet the half-time stability criteria set for this product. Preliminary investigation indicates that kit batch 1001424181 lot 27816 of the same product may also be impacted. Customers are instructed to: 1. Dispose of GXSACOMP-10 lots 27815 and/or 27816 you have in inventory. Cepheid will provide replacement product. The replacement product should ship within 5 business days following receipt of the completed response form. 2. If you have previously received or receive a negative test result with GXSACOMP-10 lots 27815 or 27816, and the patient shows clinical signs/symptoms of SA or MRSA, re-test with a cartridge from a different lot or use an alternative test/method for SA or MRSA, if available. 3. With all questionable positive or questionable negative results, please continue to report to Cepheid Technical Support for investigation. 4. Cepheid asks that you acknowledge receipt of this Urgent Medical Device Recall by completing the attached Customer Response Form and either emailing the completed form to CFQ@cepheid.com or faxing it to +1 (408) 716-3143. Cepheid will not send a reply or confirmation of received response forms.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.