CHANGE HEALTHCARE CANADA COMPANY McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management
Brand
CHANGE HEALTHCARE CANADA COMPANY
Lot Codes / Batch Numbers
McKesson Cardiology Software Versions: 13.2.1 HF1, 13.2.1 HF2, 14.0, 14.0 HF1, 14.0 HF2, Change Healthcare Software Versions: 14.1.1, 14.2
Products Sold
McKesson Cardiology Software Versions: 13.2.1 HF1, 13.2.1 HF2, 14.0, 14.0 HF1, 14.0 HF2; Change Healthcare Software Versions: 14.1.1, 14.2
CHANGE HEALTHCARE CANADA COMPANY is recalling McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management due to Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format
Recommended Action
Per FDA guidance
Change Healthcare Canada Company notified customers on about 02/10/2020 via "Urgent Field Safety Notice" letter. Customers were informed of an issue whereby users who are estimating heart rate by counting grid squares, may assume that the Sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value. Instructions included that the Simultaneous ECG format can be used rather than the Sequential ECG format, until the software is updated and the Sequential ECG format option can be used while maintaining awareness to the scale bar on the screen. Customers were also instructed to contact their users and train them on how to verify the ECG format mode they are working on and how they can change it. The notice needs to be distributed to all those within their organization who need to be aware and to please alert other parties affiliated with your organization that are affected by this field safety notice. Additionally, customers were informed that Change Healthcare is developing a software update that will enable presentation of the Sequential ECG format view using the 25 mm/sec scale and were requested to contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice and 2) Arrange installation of software update/upgrade. Change Healthcare Customer Support contacts: - US / Canada: 1 866-777-0202 - UK / Ireland: +44 (0) 208 9527399
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.