Chattanooga Corp Inc VECTRA TRANSPORTABLE, Model No. 2783, Chattanooga Group, Hixson, TN 37343, Made in the USA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VECTRA TRANSPORTABLE, Model No. 2783, Chattanooga Group, Hixson, TN 37343, Made in the USA
Brand
Chattanooga Corp Inc
Lot Codes / Batch Numbers
Serial numbers 1000 - 1615
Products Sold
Serial numbers 1000 - 1615
Chattanooga Corp Inc is recalling VECTRA TRANSPORTABLE, Model No. 2783, Chattanooga Group, Hixson, TN 37343, Made in the USA due to Device malfunction: causing electric shock and burn.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Device malfunction: causing electric shock and burn.
Recommended Action
Per FDA guidance
Beginning on 12/17/2007, Chattanooga Corp. sent two recall notification letters, via UPS mail, to its consignees informing them of the situation. The letters stated that the recall was to the user level and requested that the user perform the software upgrade, which will eliminate the possibility of shock and burn.