Churchill Medical Systems, Inc. Churchill Medical Systems Bag Access Device w/Smartsite Needleless Injection Site, Sterile, Non pyrogenic, Latex Free Product Code: KSA-140. Churchill Medical Systems, Inc. Dover, NH 03820. Intravascular administration set. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Churchill Medical Systems Bag Access Device w/Smartsite Needleless Injection Site, Sterile, Non pyrogenic, Latex Free Product Code: KSA-140. Churchill Medical Systems, Inc. Dover, NH 03820. Intravascular administration set.
Brand
Churchill Medical Systems, Inc.
Lot Codes / Batch Numbers
Lot Numbers: 903005, 904057, 904060, 904061, 904062, 904063, 905089, 905090, 905091, 906199, 907005, 907006, 907007, 909014, 909015, 909016, 909017, 909018, 909026, 909031, 909032, 909033, 909034, 909035, 909036 and 909037.
Products Sold
Lot Numbers: 903005, 904057, 904060, 904061, 904062, 904063, 905089, 905090, 905091, 906199, 907005, 907006, 907007, 909014, 909015, 909016, 909017, 909018, 909026, 909031, 909032, 909033, 909034, 909035, 909036 and 909037.
Churchill Medical Systems, Inc. is recalling Churchill Medical Systems Bag Access Device w/Smartsite Needleless Injection Site, Sterile, Non pyro due to Bond between the spike and needle may leak, break or detach resulting in leakage or spillage of chemotherapeutic agents. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Bond between the spike and needle may leak, break or detach resulting in leakage or spillage of chemotherapeutic agents
Recommended Action
Per FDA guidance
Churchill Medical Systems issued an "Urgent: Product Advisory Customer Notification" dated October 20, 2009 requesting return of affected product and notification to end users. For further information, contact Churchill Medical Systems at 1-603-743-5988 or via fax at 1-603-743-6328.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
FL