Cincinnati Sub-Zero Products Inc FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient warming). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient warming).
Brand
Cincinnati Sub-Zero Products Inc
Lot Codes / Batch Numbers
Product # 82542, Catalog # 542, Lot #: 07279, 07280, 07424 & 08813.
Products Sold
Product # 82542; Catalog # 542; Lot #: 07279, 07280, 07424 & 08813.
Cincinnati Sub-Zero Products Inc is recalling FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for pati due to Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.
Recommended Action
Per FDA guidance
On November 14, 2009, Cincinnati Sub-zero Products, Inc. (CSZ) sent a recall notification letter and response letter to customers. A second notice dated December 10, 2009 was also issued to customers. Customers were instructed in the notifications to assume the recalled product is not sterile and do not use. CSZ requested all affected product be returned to their firm for replacement. Customers are to complete and return the response letter acknowledging receipt of the recall letter. Direct questions to CSZ by calling Customer Service at 1-800- 989-7373. Customers may also call 1-800-989-7373, extension 3234 or 1-51 3-71 9-3234 for recall instructions. Representatives will be available to take your call from 8:30 a.m. until 4:30 p.m. eastern standard time, Monday through Friday.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026