Coloplast Corp Skin Care Div Coloplast, InterDry Ag 10" x 144" Roll P/N 6791001400, Manufactured for Coloplast A/S DK-3050 Humlebaek, Denmark, Distributed by Coloplast Corp Minneapolis MN 55411. InterDry" is a non-sterile, knitted, polyurethane-coated polyester textile impregnated with an antimicrobial silver complex used for the management of skin fold complications. The textile provides moisture transportation to keep skin dry while the antimicrobial in the textile reduces odor. The textile's low friction surface ac Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Coloplast, InterDry Ag 10" x 144" Roll P/N 6791001400, Manufactured for Coloplast A/S DK-3050 Humlebaek, Denmark, Distributed by Coloplast Corp Minneapolis MN 55411. InterDry" is a non-sterile, knitted, polyurethane-coated polyester textile impregnated with an antimicrobial silver complex used for the management of skin fold complications. The textile provides moisture transportation to keep skin dry while the antimicrobial in the textile reduces odor. The textile's low friction surface ac
Brand
Coloplast Corp Skin Care Div
Lot Codes / Batch Numbers
Lot 1789435
Products Sold
Lot 1789435
Coloplast Corp Skin Care Div is recalling Coloplast, InterDry Ag 10" x 144" Roll P/N 6791001400, Manufactured for Coloplast A/S DK-3050 Humle due to Coloplast determined that a portion of one lot of InterDry" Ag Lot number 1789435, did not meet the product specification for silver content. In this. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Coloplast determined that a portion of one lot of InterDry" Ag Lot number 1789435, did not meet the product specification for silver content. In this portion of the lot antimicrobial properties of the product could be compromised although the absorption and transportation of moisture would be as expected.
Recommended Action
Per FDA guidance
Consignees were faxed on 7/27/09 a Coloplast "Product Recall" letter dated July 2009. The letter addressed the problem and requested consignees to contact Coloplast Customer Service at 1-800-533-0464 to communicate the status of the product. Replacement product will be provided upon receipt of returned product. It also advised distributors to contact their customers immediately and advise them of the recall and to return the product them. Questions or concerns should be directed to the customer's customer care rep at 1-800-533-0464.