Computerized Medical Systems Inc Interplant Radiation Treatment Planning System, Interplant Release 3.4.0 Used when planning and performing prostate brachytherapy implant procedure. It provides ultrasound probe alignment and radioactive seed implantation. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Interplant Radiation Treatment Planning System, Interplant Release 3.4.0 Used when planning and performing prostate brachytherapy implant procedure. It provides ultrasound probe alignment and radioactive seed implantation.
Brand
Computerized Medical Systems Inc
Lot Codes / Batch Numbers
Interplant Release 3.4.0
Products Sold
Interplant Release 3.4.0
Computerized Medical Systems Inc is recalling Interplant Radiation Treatment Planning System, Interplant Release 3.4.0 Used when planning and per due to Interplant Software: If the trajectory needle with a non-zero retraction is adjusted in any of the planning views (Transverse, 3D) and the program is . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Interplant Software: If the trajectory needle with a non-zero retraction is adjusted in any of the planning views (Transverse, 3D) and the program is closed and reopened, the Needle Definitions page in the Dose Plan Summary (Plan View) displays the needle with a 0.00 retraction even when the needle is displayed correctly in the planning view. If the Eliminate Leading Spacers function is applied t
Recommended Action
Per FDA guidance
Computerized Medical Systems, Inc. sent a "CUSTOMER ADVISORY" letter dated December 1, 2008, to all affected customers. The letter identified the product, the problem, and the actions to be taken. Customers were instructed to discontinue use of the product. There is no workaround at this time. This issue was resolved in Interplant Release 3.4.1 software. All possibly affected users have been shipped the Interplant Release 3.4.1 software in May 2010. If you have any questions, please call 314-993-0003 or toll free 800-878-4267.