Conmed Corporation ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Spatula; Cat. #60-5158-234 -''J'' Hook; Cat. #60-5158-432 - ''L'' Hook; Cat. # 60-5158-934 - Needle. FIRM ON LABEL: ConMed Corporation, Utica, New York, USA 13502. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Spatula; Cat. #60-5158-234 -''J'' Hook; Cat. #60-5158-432 - ''L'' Hook; Cat. # 60-5158-934 - Needle. FIRM ON LABEL: ConMed Corporation, Utica, New York, USA 13502.
Brand
Conmed Corporation
Lot Codes / Batch Numbers
Cat. #60-5158-034 with lot #s 0310301, 0402031, 0403051, 0404131, Cat. # 60-5158-234 with lot # 0310081, Cat. #60-5158-432 with lot #s 0310081, 0310301, 0311101, 0401051, 0401121, 0402031, 0403051, 0411121, Cat. # 60-5158-934 with lot #s 0311101, 0401121.
Products Sold
Cat. #60-5158-034 with lot #s 0310301, 0402031, 0403051, 0404131; Cat. # 60-5158-234 with lot # 0310081; Cat. #60-5158-432 with lot #s 0310081, 0310301, 0311101, 0401051, 0401121, 0402031, 0403051, 0411121; Cat. # 60-5158-934 with lot #s 0311101, 0401121.
Conmed Corporation is recalling ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Sp due to Insulation of some electrodes may have pulled away from the hub of the electrode exposing the conductive shaft below. There is potential for electric. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Insulation of some electrodes may have pulled away from the hub of the electrode exposing the conductive shaft below. There is potential for electrical shock.
Recommended Action
Per FDA guidance
Letters dated 6/28/04 via Priority Mail with instructions to return product and to subrecall if further distributed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026