Conmed Corporation ConMed Stealth Coated Laparoscopic Electrodes with 3/32'' pin (60-5158 series) as follows: Cat. #60-5158-027 - Spatula 5 mm x 27 cm; Cat #60-5158-032 - Spatula 5 mm x 32 cm; Cat. #60-5158-044 - Spatula 5 mm x 44 cm; Cat #60-5158-127 - L Hook 5 mm x 27 cm; Cat #60-5158-132 - L Hook, 5 mm x 32 cm; Cat #60-5158-144 - L Hook 5 mm x 44 cm; Cat #60-5158-232 - J Hook 5mm x 32 cm; Cat. #60-5158-244 - J Hook 5mm x 44 cm; Cat #60-5158-927 - Needle 5 mm x 27 cm; Cat #60-5158-932 - Needle 5 mm x Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ConMed Stealth Coated Laparoscopic Electrodes with 3/32'' pin (60-5158 series) as follows: Cat. #60-5158-027 - Spatula 5 mm x 27 cm; Cat #60-5158-032 - Spatula 5 mm x 32 cm; Cat. #60-5158-044 - Spatula 5 mm x 44 cm; Cat #60-5158-127 - L Hook 5 mm x 27 cm; Cat #60-5158-132 - L Hook, 5 mm x 32 cm; Cat #60-5158-144 - L Hook 5 mm x 44 cm; Cat #60-5158-232 - J Hook 5mm x 32 cm; Cat. #60-5158-244 - J Hook 5mm x 44 cm; Cat #60-5158-927 - Needle 5 mm x 27 cm; Cat #60-5158-932 - Needle 5 mm x
Brand
Conmed Corporation
Lot Codes / Batch Numbers
Cat #60-5158-027, Lot #s 0309121, 0310301, 0401051, 0404271, 0405031, Cat #60-5158-032, Lot #s 0309011, 0309111, 0401121, 0402031, 0402271, 0404131, 0405031, 0406071, Cat #60-5158-044, Lot #s 0309151, 0403121, Cat #s 60-5158-127, Lot #s 0309021, 0309151, 0310301, Cat #60-5158-132, Lot #s 0308071, 0308211, 0310081, 0311031, 0311101, 0401051, 0402031, 0402271, 0403121, 0404131, Cat #60-5158-144, Lot #s 0308261, 0310061, 0310081, 0311101, 0402031, 0403031, 0403121, 0406071, Cat #s 60-5158-232, Lot #s 0309121, 0310301, 0402181, 0404301, Cat #60-5158-244, Lot #s 0309121, 0310081, 0402271, 0404301, Cat #60-5158-927, Lot # 0309021, Cat # 60-5158-932, Lot #0308111, 0309021, 0310081, 0311101, 0401121, 0402181, 0403121.
Products Sold
Cat #60-5158-027, Lot #s 0309121, 0310301, 0401051, 0404271, 0405031; Cat #60-5158-032, Lot #s 0309011, 0309111, 0401121, 0402031, 0402271, 0404131, 0405031, 0406071; Cat #60-5158-044, Lot #s 0309151, 0403121; Cat #s 60-5158-127, Lot #s 0309021, 0309151, 0310301; Cat #60-5158-132, Lot #s 0308071, 0308211, 0310081, 0311031, 0311101, 0401051, 0402031, 0402271, 0403121, 0404131; Cat #60-5158-144, Lot #s 0308261, 0310061, 0310081, 0311101, 0402031, 0403031, 0403121, 0406071; Cat #s 60-5158-232, Lot #s 0309121, 0310301, 0402181, 0404301; Cat #60-5158-244, Lot #s 0309121, 0310081, 0402271, 0404301; Cat #60-5158-927, Lot # 0309021; Cat # 60-5158-932, Lot #0308111, 0309021, 0310081, 0311101, 0401121, 0402181, 0403121.
Conmed Corporation is recalling ConMed Stealth Coated Laparoscopic Electrodes with 3/32'' pin (60-5158 series) as follows: Cat. #60 due to Insulation may pull away from the hub of the electrode exposing the conductive shaft below. This condition results in potential for electrical shock.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Insulation may pull away from the hub of the electrode exposing the conductive shaft below. This condition results in potential for electrical shock.
Recommended Action
Per FDA guidance
Letters dated 8/9/04 via Priority Mail with instructions to return product and subrecall if further distributed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution