Conmed Electrosurgery Conmed System 2500 Electrosurgical Unit, REF 60-8011-SYS, Conmed Corporation Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Conmed System 2500 Electrosurgical Unit, REF 60-8011-SYS, Conmed Corporation
Brand
Conmed Electrosurgery
Lot Codes / Batch Numbers
Serial numbers: 06GGN 001 thru 009, 06FGN 001 thru 010.
Products Sold
Serial numbers: 06GGN 001 thru 009, 06FGN 001 thru 010.
Conmed Electrosurgery is recalling Conmed System 2500 Electrosurgical Unit, REF 60-8011-SYS, Conmed Corporation due to Wrong size capacitor installed in electrosurgical generator could malfunction, causing muscle stimulation or delay in therapy.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Wrong size capacitor installed in electrosurgical generator could malfunction, causing muscle stimulation or delay in therapy.
Recommended Action
Per FDA guidance
Domestic customers were notified by letter on 09/01/2006 and international accounts were notified by letter on 09/07/2006.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026