Cook Inc. Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.
Brand
Cook Inc.
Lot Codes / Batch Numbers
lot 13846597
Products Sold
lot 13846597
Cook Inc. is recalling Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G due to Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
Recommended Action
Per FDA guidance
On September 28, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers should take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return to the firm. 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Cook Medical is removing potentially impacted devices from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution