Cook Inc. Pressure Monitoring Set, C-PMS, G02838 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Pressure Monitoring Set, C-PMS, G02838
Brand
Cook Inc.
Lot Codes / Batch Numbers
Lot Numbers: 8910219, 8937938, 8951879, 8979177, 8990088, 9007566, 9008052, 9039794, 9351985, 9351986, 9354853, 9359896, 9367663, 9372621, 9380316, 9381545, 9386056, 9386351, 9389523, 9395471, 9395682, 9395683, 9402026, 9403779, 9406107, 9409058, 9412543, 9414585, 9416780, 9486988, 9785433, 9799632, 9799638, 9826828, 9826829, 9829654, 9830524, 9836105, 9836209, 9871734, 9873996, 9880405, 9889391, 9892371, 9898730, 9910485, 9916495, 9916503, 9924121, 9931357, 9931541, 9938235, 9946241, 9952626, 9952627, 9952726, 9954044, 9954213, 9961061, 9969348, 9975178, 10109034, 10111785, 10130377, 10130378, 10134000, 10148417, NS8927417
Products Sold
Lot Numbers: 8910219, 8937938, 8951879, 8979177, 8990088, 9007566, 9008052, 9039794, 9351985, 9351986, 9354853, 9359896, 9367663, 9372621, 9380316, 9381545, 9386056, 9386351, 9389523, 9395471, 9395682, 9395683, 9402026, 9403779, 9406107, 9409058, 9412543, 9414585, 9416780, 9486988, 9785433, 9799632, 9799638, 9826828, 9826829, 9829654, 9830524, 9836105, 9836209, 9871734, 9873996, 9880405, 9889391, 9892371, 9898730, 9910485, 9916495, 9916503, 9924121, 9931357, 9931541, 9938235, 9946241, 9952626, 9952627, 9952726, 9954044, 9954213, 9961061, 9969348, 9975178, 10109034, 10111785, 10130377, 10130378, 10134000, 10148417, NS8927417
Cook Inc. is recalling Pressure Monitoring Set, C-PMS, G02838 due to The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occl. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
Recommended Action
Per FDA guidance
On April 30, 2020, Cook Medical sent an Urgent Medical Device Recall letter to their customers to inform them of the voluntary recall of specific lots of the Pressure Monitoring Set and Pressure Monitoring Tray. The firm instructed customers to examine their inventory immediately to determine if have affected product(s) and quarantine any affected product that remains unused then, cease all distribution and use of the product. Cook Medical requested that customers complete the Acknowledgement and Receipt Form via fax or email within 5 business days of receiving the recall letter, whether or not they had any affected product(s) on hand. Customers were instructed to return the affected product(s) with a copy of the Acknowledgement and Receipt Form to receive a product credit. ***Updated 7/7/20*** The firm sent an updated letter dated July 7, 2020, to inform customers of 14 additional lots of the Pressure Monitoring Tray (GPN G02854) that were inadvertently excluded from the previous letter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, DC