Cook Incorporated Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476
Brand
Cook Incorporated
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
THSCF-25-180-3 G02165 UDI-DI: 00827002021655 Lots/Expiration Dates: 14797761 16-06-2027 14813704 27-06-2027 TSCF-25-80-3 G00476 UDI-DI: 00827002004764 Lots/Expiration Dates: NS14797766 16-06-2027 NS14807600 22-06-2027 NS14821727 30-06-2027
Cook Incorporated is recalling Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476 due to Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the ste. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Recommended Action
Per FDA guidance
On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235