Cook Incorporated Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498) Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498)
Brand
Cook Incorporated
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lots: 14212811 14215685 14212247 14212812 14212814 14215682 14361413 14361416 14285326 14285328 14285330 14297289 14361412 14445407 14444454 14477949 14477950 UDI: (01)00827002474994(17)261013(10)14212811 (01)00827002474994(17)261008(10)14215685 (01)00827002474987(17)261011(10)14212247 (01)00827002474987(17)261008(10)14212812 (01)00827002474994(17)261008(10)14212814 (01)00827002474994(17)261008(10)14215682 (01)00827002474994(17)261219(10)14361413 (01)00827002474994(17)261220(10)14361416 (01)00827002474987(17)261110(10)14285326 (01)00827002474987(17)261110(10)14285328 (01)00827002474994(17)261110(10)14285330 (01)00827002474994(17)261123(10)14297289 (01)00827002474987(17)261220(10)14361412 (01)00827002474987(17)270207(10)14445407 (01)00827002474994(17)270207(10)14444454 (01)00827002474987(17)270217(10)14477949 (01)00827002474994(17)270216(10)14477950
Cook Incorporated is recalling Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (G due to The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.
Recommended Action
Per FDA guidance
On May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user. Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate personnel, down to the user level, within your organization or with any organization where the affected devices may have been transferred. Physicians should practice their standard patient monitoring following the procedure for an early identification of any complications to mitigate their severity. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.