CooperSurgical, Inc. TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or betwee Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or betwee
Brand
CooperSurgical, Inc.
Lot Codes / Batch Numbers
IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390
Products Sold
IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390
CooperSurgical, Inc. is recalling TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer W due to The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is pro. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.
Recommended Action
Per FDA guidance
On July 8, 2019, the firm began notifying customers of the recall via Urgent Medical Device Correction letters. The letters informed customers that CooperSurgical has updated the IFU to clarify that use of the Transwarmer Mattress with other heat producing devices, such as an incubator, is prohibited. Such use could lead to serious health consequences, such as skin burns. Customers were provided with an updated version of the IFU. Customers were asked to complete and return the acknowledgement and receipt form to the firm. If you have any questions about this recall, please contact CooperSurgical at 203-601-5200 ext. 3300.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026