Cordis Corporation CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease.
Brand
Cordis Corporation
Lot Codes / Batch Numbers
10 lots. Lot #s: 13326414*, 13326415*, 13327582, 13329185*, 13332369, 13333077*, 13339105*, 13339106*, 13341244, 13343464. (*These six lots have U.S. distribution.)
Products Sold
10 lots. Lot #s: 13326414*, 13326415*, 13327582, 13329185*, 13332369, 13333077*, 13339105*, 13339106*, 13341244, 13343464. (*These six lots have U.S. distribution.)
Cordis Corporation is recalling CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. I due to Data transcription error. The 10 affected lots of Catalog Number C10040SL were manufactured with a prior version of the guidewire lumen material, whi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Data transcription error. The 10 affected lots of Catalog Number C10040SL were manufactured with a prior version of the guidewire lumen material, which is not validated with the current manufacturing process. This condition may result in a separation of the device. If a separation was to occur, and the separated fragment of lumen material was not retrieved on the guidewire, blockage of the vessel
Recommended Action
Per FDA guidance
A letter (RE: Medical Device Recall) dated March 11, 2008, and Acknowledgement Form was sent overnight to the following 6 individuals in each account: Dir. Cardiac Cath Lab, Dir. Radiology Services, Dir. Vascular Services, Risk Manager, Materials Dir., Hospital Administrator; requesting that the affected product be set aside, and that the form be reviewed, signed, and faxed back to Cordis. Affected product is to be either returned or the local sales rep should the contacted to facilitate the return of the affected product. All product will be replaced. Customer Service can be contacted if there are additional questions relative to product return, replacement, or availability.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.