Covidien Llc Puritan Bennett 560 Ventilator, PB560, CFN 4096600 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Puritan Bennett 560 Ventilator, PB560, CFN 4096600
Brand
Covidien Llc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
UDI-DI: 10884521196377 U.S. Serial Numbers: 40966N7229 40966N7231 40966N7334 40966N7335 40966N7340
Covidien Llc is recalling Puritan Bennett 560 Ventilator, PB560, CFN 4096600 due to Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufactu. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
Recommended Action
Per FDA guidance
Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures. All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to replace the affected ventilator turbine component. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution