Covidien LLC SIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT
Brand
Covidien LLC
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
N6E0033UX N6H0768UX N6H0772UX N6K0521UX N6K0522UX N6K0667UX N6K0668UX N6L0687UX N6M0399UX N7A0018UX N7A0213UX N7A0401UX N7A0837UX N7B0787UX N7C0238X N7C0470X N7E0496X N7E1088X N7F0670X N7H0962X N7J0370X N8A0188X N8B0057X N8B0483X N8B0940X N8D0365X N8D0649X N8E0092X N8E0213X N8E0516X N8E1178X N8F0483X N8F0819X N8F0974X N8J0440X N8L0589Y N8M0529Y N9A0050Y N9A0427Y
Covidien LLC is recalling SIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT due to The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lea. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Recommended Action
Per FDA guidance
All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026