Cygnus LLC Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
Brand
Cygnus LLC
Lot Codes / Batch Numbers
Serial Numbers: For MTF: -- 034135 through 034144, For Cascade -- 034119 through 034128.
Products Sold
Serial Numbers: For MTF: -- 034135 through 034144; For Cascade -- 034119 through 034128.
Cygnus LLC is recalling Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch S due to The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies th. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
Recommended Action
Per FDA guidance
A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
NY