Cytocell Ltd. 418 Cambridge Science Park, Milton Road Oxford Gene Technology Cambridge United Kingdom Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024
Brand
Cytocell Ltd. 418 Cambridge Science Park, Milton Road Oxford Gene Technology Cambridge United Kingdom
Lot Codes / Batch Numbers
Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711
Products Sold
Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711
Cytocell Ltd. 418 Cambridge Science Park, Milton Road Oxford Gene Technology Cambridge United Kingdom is recalling Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from due to individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. T. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Recommended Action
Per FDA guidance
Cytocell issued Urgent Medical Recall Reference: VC/2022/001 letter on 1/18/22 to Distributors and End Users via email. Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of the probe being manufactured with the fluorophores the incorrect way around. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.