Daavlin Distributing Company 3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424.
Brand
Daavlin Distributing Company
Lot Codes / Batch Numbers
All Touch Control System, Model #PASNBSM-2424 units manufactured between 6/5/03 and 4/30/04.
Products Sold
All Touch Control System, Model #PASNBSM-2424 units manufactured between 6/5/03 and 4/30/04.
Daavlin Distributing Company is recalling 3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424. due to Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
Recommended Action
Per FDA guidance
The firm has software patches to correct the programming problems as they are reported by the users. These patches are being installed either by a service representative for the manufacturer or by the customer
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026