Datex Ohmeda, Inc GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The EZchange Module is an optional accessory for these anesthesia systems. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The EZchange Module is an optional accessory for these anesthesia systems.
Brand
Datex Ohmeda, Inc
Lot Codes / Batch Numbers
EZchange Manifold with part number 1407-7021-000, 1407-7027-000, M1053007, M1053008.
Products Sold
EZchange Manifold with part number 1407-7021-000, 1407-7027-000, M1053007, M1053008.
Datex Ohmeda, Inc is recalling GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Ane due to Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
Recommended Action
Per FDA guidance
Because this issue only occurs when the Disposable Multi Absorber is used in conjunction with EZchange Module, they have contacted all EZchange Module customers. It is critical to contact these customers specifically, rather than all customers shipped Disposable Multi Absorbers as this may cause unwarranted confusion with customers who do not have the EZchange Module. Consignees were sent an letter dated 7/19/2007 describing the product and problem. The letter lists a number of interim actions to implement with the anesthesia systems.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026