Del Medical Systems Group Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column Extensions; a radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Del Medical Systems Group, 11550 West King Street, Franklin Park, IL 60131-1330. The device is used for X-Ray tube positioning. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column Extensions; a radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Del Medical Systems Group, 11550 West King Street, Franklin Park, IL 60131-1330. The device is used for X-Ray tube positioning.
Brand
Del Medical Systems Group
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
All Model CTM and DCTM Ceiling Tube Mounts with either a 5 inch extension (Part 114-0304G2) or 10 inch extension (Part 114-0304G1). There are no serial numbers on the extensions.
Del Medical Systems Group is recalling Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch due to There is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or oper. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator.
Recommended Action
Per FDA guidance
Del Medical sent a "Field Correction Recall" letter with an acknowledgement form to the direct account dealers/assemblers on January 12, 2009 - February 4, 2009, informing them that required maintenance inspection of a retaining ring inside the column assembly may not be performed as required. If the retaining ring is not regularly inspected and replaced when worn or loose, there is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator. The accounts were requested to locate the affected systems and immediately take them out of service until correction by installing a Cable Mount Retention Retrofit Kit to eliminate the potential for the cable to become detached. Affected accounts were instructed to complete the enclosed acknowledgement form and return it via fax (847-288-7911) or mail to Del Medical. If there are any questions, please contact Del Medical Technical Support at 800-800-6006 or 847-288-7000.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.