DePuy Orthopaedics, Inc. RECLAIM Distal Reamer Extension Non Sterile REF 2975-00-500 Intended as an attachment device to the distal reamers to allow the surgeon to clear soft tissue while reaming the distal femoral canal. Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
RECLAIM Distal Reamer Extension Non Sterile REF 2975-00-500 Intended as an attachment device to the distal reamers to allow the surgeon to clear soft tissue while reaming the distal femoral canal.
Brand
DePuy Orthopaedics, Inc.
Lot Codes / Batch Numbers
J0111, J0211, J0411, J0611, J0911, J1211, NB12649, NB12650, NB12651, NB12652, NB12653, NB12654, NB12655, NB12656, NB23480, NB23481, NB3584, NB3586, NB3587, NB3588, NB3590, NB3591, NB5955, NB7894, NB8016, NB8017, NB8018, NB8019, NB8020, NB8021, NB8022, NB8023, 578910, 578913, R578910, R578912, R578913, 582631, and 592125.
Products Sold
J0111, J0211, J0411, J0611, J0911, J1211, NB12649, NB12650, NB12651, NB12652, NB12653, NB12654, NB12655, NB12656, NB23480, NB23481, NB3584, NB3586, NB3587, NB3588, NB3590, NB3591, NB5955, NB7894, NB8016, NB8017, NB8018, NB8019, NB8020, NB8021, NB8022, NB8023, 578910, 578913, R578910, R578912, R578913, 582631, and 592125.
DePuy Orthopaedics, Inc. is recalling RECLAIM Distal Reamer Extension Non Sterile REF 2975-00-500 Intended as an attachment device t due to Specific lots are being recalled due to the potential for the tabs to break, leaving fragments in the patient.. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Specific lots are being recalled due to the potential for the tabs to break, leaving fragments in the patient.
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.