DePuy Orthopaedics, Inc. Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006). Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).
Brand
DePuy Orthopaedics, Inc.
Lot Codes / Batch Numbers
CATALOG NO. : ATTUNE INTUITION Impactors Rotating Tibial Platform, (254401004) Barcode GTIN Number: Rotating Tibial Platform, ......... 10603295130208 ALL LOTS
Products Sold
CATALOG NO. : ATTUNE INTUITION Impactors Rotating Tibial Platform, (254401004) Barcode GTIN Number: Rotating Tibial Platform, ......... 10603295130208 ALL LOTS
DePuy Orthopaedics, Inc. is recalling Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable i due to ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be o. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026