DeRoyal Industries Inc DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Brand
DeRoyal Industries Inc
Lot Codes / Batch Numbers
GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Products Sold
GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
DeRoyal Industries Inc is recalling DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B due to The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Recommended Action
Per FDA guidance
The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. If you have questions regarding this recall please contact us at 865-362-6465 or email us at recalls@deroyal.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026