DiaDexus, Inc PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit, manufactured by DiaDexus, South San Francisco, CA. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit, manufactured by DiaDexus, South San Francisco, CA.
Brand
DiaDexus, Inc
Lot Codes / Batch Numbers
Lot numbers 078011, 088061, 098091, 109011, 119021, 129031, 139051, 139051A
Products Sold
Lot numbers 078011, 088061, 098091, 109011, 119021, 129031, 139051, 139051A
DiaDexus, Inc is recalling PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia t due to Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be subjected to additional lifestyle changes or intensified pharmaceutical therapy; such as statins, as a result of the falsely elevated result.
Recommended Action
Per FDA guidance
Notification letter issued to consignees on 10/29/2009. Customers were instructed to not report PLAC test results for samples that are greater than 360 ng/mL, and pursue confirmatory testing by contacting DiaDexus. Customers with questions please call the company technical support line toll free at 1-877-752-2837 for assistance. UPDATE: On April 30, the firm suspended marketing of the PLAC TIA method, as the limited range no longer covered the clinical data/reference range data cited in the product insert. The firm sent out a letter notifying customers on May 10, 2010. The firm set up a sample send out program, so that customers can send samples to labs running the ELISA method PLAC test. Arrangements were made so that labs could be enabled to run a CLIA complex assay, for customer credit programs for un-used products, and a fax back response form for response tracking.