Diagnostic Chemicals, Ltd. Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892
Brand
Diagnostic Chemicals, Ltd.
Lot Codes / Batch Numbers
Lot 29892
Products Sold
Lot 29892
Diagnostic Chemicals, Ltd. is recalling Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892 due to Upon subsequent testing on different instruments (as required by other customers), DCL detected a non-linear response on certain analyzers. A decision. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Upon subsequent testing on different instruments (as required by other customers), DCL detected a non-linear response on certain analyzers. A decision was made to recall the product due to the fact that DCL customers sell the product to different end users, DCL cannot be certain which instruments the end users employ in their laboratories
Recommended Action
Per FDA guidance
For product shipped by DCL to one customer notification letter dated 12/5/06 sent out to customer to return/discard product. Replacement will be sent to customer. For product shipped to DCL (USA) warehouse notification sent to discard kits remaining in stock. For product remaining at DCL Canada in-house notification sent out to discard in-house kits.