Diasorin Inc. Parvovirus B19 IgG EIA kit Part No. V519IGUS Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Parvovirus B19 IgG EIA kit Part No. V519IGUS
Brand
Diasorin Inc.
Lot Codes / Batch Numbers
Lot 71440040AA
Products Sold
Lot 71440040AA
Diasorin Inc. is recalling Parvovirus B19 IgG EIA kit Part No. V519IGUS due to Printed label inside the box lid of the Parvovirus has the incorrect Lot Specific Constant value. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Printed label inside the box lid of the Parvovirus has the incorrect Lot Specific Constant value
Recommended Action
Per FDA guidance
DiaSorin issued a customer notification letter on July 17, 2020. The letter states the following action to be taken: Please use the Lot Specific Constant value of 0.14 with your remaining inventory of Lot 71440040AA.. The Lot Specific Constant is used to calculate the cut-off value and use of the incorrect value may lead to: 1. An invalid test due to the Low Positive Control resulting above the upper limit of the assigned range. 2. The reporting of a false positive patient result if the Low Positive Control does not fail and invalidate the test run. If you have used the affected product Lot 71440040AA in combination with the incorrect Lot Specific Constant of 0.10 and you obtained valid runs, please: 1. Review the test results of these valid runs for values between 1.1 Index and 1.54 Index, which may have been falsely reported as positive. 2. Take actions as appropriate. 3. DiaSorin recommends discussing the content of this letter with your laboratory director with respect to the need to review previous test results, conduct patient follow-up and/or repeat testing as appropriate. Complete page two of this letter and return to DiaSorin Inc. Forward this letter to anyone in your organization who may be impacted. We apologize for any inconvenience this may cause. If you have any questions regarding this notice please contact DiaSorin Inc. Technical Support at 1-800-328-1482.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, CT, IL, MA, MI, MN, NJ, NY, NC, VA, WI