DPC Cirrus IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836
Brand
DPC Cirrus
Lot Codes / Batch Numbers
All analyzers, all serial numbers.
Products Sold
All analyzers; all serial numbers.
DPC Cirrus is recalling IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufa due to The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.
Recommended Action
Per FDA guidance
The firm became aware of the problem on 4/19 and conducted tests on 4/22/05. A service alert for field service engineers and Immulite 2000 distributors was prepared on 4/22/05.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.